WCaution
To avoid damaging fiber connectors, to avoid data
loss, and to ensure maximum accuracy of test results:
•
Use proper cleaning procedures to clean all fiber
connectors before every use. Neglecting this step or
using improper procedures can cause unreliable test
results and may permanently damage the
connectors.
•
Cover all connectors with protective caps when not
in use.
•
Never connect the source to an active network.
Doing so can disrupt network operations.
•
After you turn on the source, let it warm up for 10
minutes before use. Allow additional time if the
source has been stored above or below ambient
temperature.
Features
fft01.eps
A
X: On/off key.
B
A: Selects auto wavelength mode, which is used with a
SimpliFiber Pro meter. The AUTO LED lights. The source’s
signal includes an identifier that tells the SimpliFiber Pro
meter which wavelength to measure.
Press
Z to change the wavelength. The wavelength LEDs
indicate the wavelength.