SpO2 Functional Tester
Detailed Specifications
15
With Fluke Biomedical R-curves
95 to 100 % ................................................. ±(3 counts + specified accuracy of the UUT)
85 to 90 % ................................................... ±(5 counts + specified accuracy of the UUT)
80 % ............................................................ ±(7 counts + specified accuracy of the UUT)
Heart Rate
Rates................................................................... 30, 40, 60, 65, 80, 100, 120, 140, 150, 180, and 240 BPM
Accuracy.............................................................. ±1 % of setting
Transmission (Ratio of detector current to LED current, expressed in parts per million (ppm))
Ratios .................................................................. Large finger (12.00 ppm), medium finger (80.00 ppm), and small finger (300.00 ppm)
Accuracy.............................................................. +50 %/-30 % for compatible monitors, unspecified for others. Selected by finger size and
color: large finger, medium finger, small finger.
Pulse Amplitude
Amplutudes ......................................................... Low (0.2 %), medium (2 %), and high (10 %
Artifact
Respiration
Size ................................................................. 2.5 % of transmission
Rate ................................................................. 20 BrPM
Ambient Light Frequency .................................... 50 Hz and 60 Hz
Compatible Manufacturer Products
With manufacturer R-curve ................................. Nellcor, Masimo, Nonin, and Nihon Kohden
With Fluke Biomedical R-curve ........................... Mindray, GE-Ohmeda, Philips/HP, and BCI