EasyManua.ls Logo

Fluke SPOT Light - General Specifications; Detailed Specifications; Oximeter Spo 2 Optical Emitter and Detector

Fluke SPOT Light
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPOT Light
Users Manual
14
General Specifications
Temperature
Operating ............................................................ 10 °C to 40 °C (50 °F to 104 °F)
Storage ............................................................... -20 °C to +60 °C (-4 °F to +140 °F)
Humidity ................................................................. 10 % to 90 % non-condensing
Altitude ................................................................... 3,000 m (9,843 ft)
Size (W x H x D) ..................................................... 12.53 cm x 14.86 cm x 4.77 cm (4.94 in x 5.85 in x 1.88 in)
Display ................................................................... LCD Monochrome display
Communication (USB Device Virtual COM Port) .. Mini B connector for service upload of firmware
Power ..................................................................... Lithium-Ion rechargeable, 3.7 V, 10.75 Wh battery, 2900 mAh
Battery Charger ..................................................... 100 V to 240 V, 50/60 Hz input, 6V/3A output. For best performance, battery charger
must be connected to a properly grounded ac receptacle.
Battery Life ............................................................ 10 hours (minimum)
Weight .................................................................... 0.29 kg (0.7 lb)
Safety ..................................................................... IEC 61010-1: 300 V CAT II, Pollution Degree 2
Ingress Protection Rating .................................... IEC 60529: IP20
Electromagnetic Environment ............................. IEC 61326-1: Basic
Detailed Specifications
Oximeter SpO
2
Optical Emitter and Detector
%O
2
O
2
Saturations ..................................................... 70 %, 75 %, 80 %, 85 %, 90 %, 95 %, 97 %, 98 %, 99 %, and 100 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range .......... ±(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy