Fora V30 - Error Messages

Fora V30
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
Error Messages
MESSAGE WHAT IT
MEANS
WHAT DOES
METER SAY?
WHAT TO DO
Appears when
the
batteries cannot
provide enough
power for a test.
Battery is dead.
Please replace.
Replace the bat-
teries immedi-
ately.
You may have
removed the
strip after apply-
ing blood to the
absorbent hole.
Test strip has
been removed
during testing.
Appears when a
used test strip
is inserted.
The test strip
has been used.
Please replace.
Re-test with a
new test strip.
Repeat with a
new test strip.
Appears when
ambient
temperature is
below system
operation range.
System op-
eration range is
50°F to 104°F
(10°C to 40°C).
Repeat the test
after the
meter and test
strip are in the
above tempera-
ture range.
Room tem-
perature is
out of range.
Thus unable to
measure.
Appears when
ambient
temperature is
above system
operation range.