4 USER’S INSTRUCTIONS
M-7.3-16 Rev. 05 09/03/15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 14
Use and maintenance manual
4.1 START UP OF APPLIANCE
Before switching on the unit, check that the electrical connections have been made correctly and
above all, that the ground connection is available and working properly (Sect. 2.2.). Perform
preliminary cleaning of the unit as described in Section 5.1.
SWITCH ON THE UNIT
To start the appliance, it is necessary to press button ON on the display. The main screen will be
displayed and message « FRIMED » will appear. Then you will be requested to enter the user
password 555.
USER PASSWORD = 555
TURN OFF THE APPLIANCE
Press the button OFF on the display, you will be asked to enter the user password. If password is
correct, after entering the last digit, the appliance switches off.
4.2 FUNCTIONS
ANTIPANIC FUNCTION
In all menu screens, after 120 seconds the key button has been pressed, the controller returns to
the main menu, without saving any unconfirmed changes.
LCD FUNCTION – ENERGY SAVING
One minute after pressing any key, the backlight of the screen switches off automatically (stand-by
status). Lighting will be switched on again by pressing any key.
DATA LOGGER
When the appliance is on, alarms and data of probes are recorded. Each single record is related
to date and time. It’s possible to record a maximum of 46000 entries (which corresponds to about
one month recording). Then you have to download records on a USB pen drive (a specific alarm
will advise you when memory is full).
In case you don’t want to download records, press ALARM key and then MUTE key. The alarm
message will disappear and the controller will overwrite on previous registrations.