4 USER’S INSTRUCTIONS
M-7.3-16 Rev. 05 09/03/15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 20
Use and maintenance manual
3 - Data load – if parameters need to be modified or updated, please insert USB pen and by
pressing « data load», the appliance will be automatically programmed with new data.
4.7 PROCEDURE TO DISABLE THE DATA LOGGER
- Press any key to activate the display
- Press MENU key
- By using the arrow keys go to point 4-PASSWORD+ENTER
- Select USER PASSWORD. By means of arrow keys enter password 555 (5+ENTER,
5+ENTER 5+ENTER)
- In this way you enter the MENU
- Scroll up to point 0-USER PARAMETERS+ENTER
- Then scroll up to point 3-RECORDINGS+ENTER
- By means of arrow keys change YES to NO and press ENTER to confirm
- Exit programming by means of ESC key to return to initial display screen
4.8 PROCEDURE TO CALIBRATE INTERAL TEMPERATURE PROBE S1
- Press any key to activate the display
- Press MENU key
- By using the arrow keys go to point 4-PASSWORD+ENTER
- Select USER PASSWORD. By means of arrow keys enter password 555 (5+ENTER,
5+ENTER, 5+ENTER)
- In this way you enter the MENU
- Scroll up to point 0-USER PARAMETERS+ENTER
- Then scroll up to point 5-CALIBR.PROBES1+ENTER
- At this point insert the new value and press ENTER to confirm
- Exit programming by means of ESC key to return to initial display screen
Example on how to choose the new parameter to enter: if the temperature showing on display is
+4°C, while inside the appliance with a different instrument +6°C is registered, the value to be
entered for calibration of probe will be +2°C
4.9 UPDATE OF FIRMWARE RELEASE
Check nr. of release on display
A. Press MENU
B. Scroll up to INFO (2)+ENTER
C. Scroll up to REL.FIRMWARE = (NUMBER)
2) TYPE OF PROCEDURE TO UPDATE FIRMWARE N. 0.32 OR PREVIOUS ONES
If Firmware installed on your appliance is n.032 or previous to this one, please perform the
following procedure:
A. Copy updated firmware on USB pen (DON’T OPEN THE FILE IN YOUR PC)
B. Switch off the appliance by pressing “OFF” key and by typing password (USER= 555)
C. Take power off (unplug)
D. Press at the same time the three keys from left below the display and plug again by keeping
the keys pressed.
E. As soon as display will show various writing, stop pressing the three keys.
F. Insert the USB pen in the port hole