EasyManua.ls Logo

FRIMED AF70E - Page 21

Default Icon
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 USERS INSTRUCTIONS
M-7.3-16 Rev. 05 09/03/15 BLOOD BANK REFRIGERATORS AND PLASMA FREEZER 21
Use and maintenance manual
G. When the word “OFF” will show, take off the USB pen and switch on the appliance again.
H. In some cases the word “PARAMETER ALARMcan show (depending on firmware release). In
order to reset the alarm, press the key “ALL” (third from left side below display)+MUTE key.
Repeat the operation until the alarm disappears.
At this point, double check that new Firmware release number has been installed
3) TYPE OF PROCEDURE TO UPDATE FIRMWARE N. 0.35 TO CURRENT RELEASE
If Firmware installed on your appliance is n.035 or following to this one, please perform the
following procedure:
A. Copy updated firmware on USB pen (DON’T OPEN THE FILE ON YOUR PC)
B. Switch off the appliance by pressing “OFF” key and by typing password (USER= 555)
C. Press the MENU key and scroll up to PASSWORD(4)+ ENTER
D. Type password (USER= 555)
E. Go to main menu and scroll up to “UPDATE SOFTWARE” (point n.5 for user)+ENTER
F. By means of arrows choose “YES” and confirm with ENTER.
G. Follow instructions showing on display
H. At this point various writing will show. Insert USB pen in the port hole
I. When the word “OFF” will show, take off the USB pen and switch on the appliance again.
J. In some cases the word “PARAMETER ALARMcan show (depending on FW release). In
order to reset the alarm, press the key “ALL” (third from left side below display)+MUTE key.
Repeat the operation until the alarm disappears.
At this point, double check that new Firmware release number has been installed
4.10 GSM MODULE (optional)
When an alarm is activated, a message is forwarded to pre-set mobile phone numbers
automatically, with:
Name, Serial Number, Alarms codes and description of alarm type.
Moreover, in case appliance includes back-up battery, if power supply is failing, a message is
forwarded to selected numbers indicating above mentioned information and “ON BATTERY”.
When situation has been restored, a message will follow indicating “POWER OK”.
By sending a SMS with the word “STATUS” to GSM number, installed into the appliance, you can
make an inquiry on its status. You will receive either a reply message with “NO ALARMS and
value of recorded temperature, in case no alarms are present, or the alarms codes detected.
Symbols on display when GSM is connected are:
GSM is connected
Field intensity blinking symbol indicates that no field is present
Symbol flashing in case of wrong PIN entering exit configuration/set-up, switch off
the appliance and enter PIN number (after three wrong entries, the system is blocked)
Symbol blinking it means that SIM hasn’t been inserted
General Error substitute GSM module

Related product manuals