EasyManua.ls Logo

Funke Gerber LactoFlash - Error messages; Troubleshooting; Error on the first measuring cell; Error on the second measuring cell; Error on both measuring cells

Funke Gerber LactoFlash
10 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Initial Operation / Unpacking and Checking the Scope of Delivery
The following parts must be present:
a) LactoFlash
b) 1 mains cable
c) 1l cleaning solution
d) 2 l Funke-Gerber Zero-Diluent
d) Report printer
e) 1 electricity supply cable (6 V) for report printer
f) Connection cable for the parallel interface for connecting the printer
g) Operating instructions for LactoFlash and printer
h) 1 hose for the output (waste)
i) Erlenmeyer flask, 100 ml
2. Initial Operation / Setting up
Place the device on a horizontal surface.
3. Initial Operation / Electrical Connection of the Device
Power connection: 230 V / 180 VA or 115 V / 180 VA
The power connection is located on the back side of the device together with
the on/off switch, voltage selector switch and mains fuse.
First select the correct mains voltage using the voltage selector switch: 230 V
or 115 V.
4. Initial Operation / Connecting the Power Cable
Ensure that the device is connected to a shockproof socket that has been
installed in accordance with the safety regulations.
5.
Initial Operation / Connecting the Hose
Connect the waste hose to the connector located on the return of the device
and lead the end of the hose either to a waste vessel of corresponding size
(approx. 5-10 l) or directly to a drain.
4 LactoFlash 31. Januar 2013
For milks of the same class (e.g. different full cream milks) the falsification by
entrainment is below the dissolving limit and thus insignificant.
However, if you switch to a very different product e.g. from skimmed milk to
cream or vice versa the first measurement in the new area must be discarded
as it is falsified by entrainment.
Error messages / Troubleshooting
Following error messages appears on the display and will printed:
Error message Cause / Resolution
Error
error on the first
measuring cell
1. There was not enough sample quantity in the bottle -- the
device has sucked in air.
2. Leak in the hose. Check the hose for leaks.
3. The sample quantity is too small. Due to wear of the pump
head, the specified minimum quantity can no longer be sucked in.
Check sample quantity, replace pump head if necessary.
4. Fault in the one or both measuring cells.
Please contact Funke-Gerber in order to replace.
.
Error
error on the
second
measuring cell
Error
error on both
measuring cells
Error
measurement is
not usable
Software failure: measurement value is not reliable. Please repeat
measurement.
Error
sample not
measureable
Sample too "thick". The fat or SNF content or both together is too
high. The sample must either be diluted or cannot be analysed.
LactoFlash 31. Januar 2013 17