58
SPECIFICATIONS
Measuring Method Oscillometric Measurement
Indication Digital LCD display
Measuring Range: Pressure: (30~280) mmHg
Pulse: (40~199) Beat/min
Accuracy: Static Pressure: ±3 mmHg
Pulse: ±5%
Memory: Memory for the last measurement
Power supply: 4x1.5V Batteries (LR6 or AA)
Operating condition: +5°C~+40°C. 30% RH~80% RH
Atmospheric pressure: 80 kPa~106
kPa
Storage condition: -20°C~+55°C. 10% RH~93% RH
Atmospheric pressure: 50 kPa~106
kPa
Dimensions: Not more than 138 (W) X 96 (H) X
50 (D) mm
Weight: Not more than 0,355 kg, excluding
batteries
Classication Type BF
Cuff Size 22-32 cm
* The specications and the outer design are subject to change for
improvement reasons without previous notice.
This device has been tested to meet the electrical and safety requirements of:
IEC/EN 60601-1,
IEC/EN 60601-1-2
Reference to Standards:
• DSTU EN 1060-1:2015 “Non-Invasive Sphygmomanometer.” General
Requirements. Paragraph 9.2. DSTU EN 1060-3:2015 “Non-Invasive
Sphygmomanometer.” Part 3. Paragraph 9.2.
• EN 1060-1/-3, NIBP-requirements
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC