GASSONIC OBSERVER-i
75
Total Calibration Event Counter (0x5E)
This reads the total number of calibration events that have been stored in the unit.
Clear all events (0x5F)
Writing zero (0) to this register clears all event counters.
Setting Clock
Please see the table on the following page.
Address
(hex)
Parameter Function Data Type Data range Access
30 Seconds Time Hi Seconds Time Hi Numeric value 0 – 65535
Timer Sec
31 Seconds Time
Low
Seconds Time Numeric value 0 – 65535
Timer sec
32 Real Time Clock
Year, Month
Read/Set year and
month of RTC
2 Numeric
Values
0-99 year, 1 –
12 month
Timer Struct
33 Real Time Clock
Day, Hour
Read/Set day and
hour of RTC
2 Numeric
Values
1 – 31 day, 0
– 23 hour
34 Real Time Clock
Minute, Second
Read/Set minutes
and seconds of
RTC
2 Numeric
Values
0 – 59 minute,
0 – 59 second
Timer Struct
35 PowerCycleFlag Read Power Cycle
Flag.
Numeric Value 1 – Time Not
Reset; 0 –
Time Was
Reset
Flag
36 Event Index Event index of
Logged Event
Numeric value 0 - 9
Index
37 Warn
Seconds Time Hi
Seconds Time Hi
for warning event
log entries
Numeric value 0 – 65535
Warn
38 Seconds Time
Low
Seconds Time Low
for warning event
log entries
Numeric value 0 – 65535
Warn
39 Structure time Hi Hi byte – year, low
byte – month for
warning event log
entries
Numeric value 0 – 65535
Warn
3A Structure time
Mid
Hi byte – day, low
byte – hour warning
event log entries
Numeric value 0 – 65535