Cleaning and disinfectant agent warnings
• If using disinfection wipes, make sure they have the same active ingredients in the same
concentrations listed above and include no other ingredients.
• Always use protective eyewear and gloves when cleaning and disinfecting any equipment.
• If a pre-mixed solution is used, be sure to observe the solution expiration date.
• Using non-recommended cleaning agent or disinfectants, using incorrect solution strengths, or
exposing the coil to the cleaner or disinfectant longer than recommended can damage or discolor the
coil and accessories.
• Follow the recommendations of the disinfectant and enzymatic cleaner manufacturer.
• Do not use automated cleaning and/or disinfection units. Follow the instructions within this
document for effective, manual cleaning and disinfection.
• Never use a bleach solution more than 250ppm.
• Never use bleach Wipes