EasyManua.ls Logo

GE AKTAprime plus - Troubleshooting Guide; UV Curve Problems

GE AKTAprime plus
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Troubleshooting
6.1 UV curve problems
Corrective actionPossible causeError symptom
Clean the system. Make sure air is
removed.
Dirt or residues in the
flow path from previ-
ous runs. Air in the
eluents.
Ghost peak
Clean the column according to the
column instructions.
Residue in the column
from previous runs
Check the mixer function by plac-
ing a stirrer bar on top of the mixer
housing. The stirrer bar should ro-
tate when the system is in Run
mode. The mixer function can also
be checked by running the installa-
tion test.
Incorrect mixer func-
tion
Clean the UV cell by flushing
Decon™ 90, Deconex™ 11 or
equivalent.
Dirty UV cellNoisy UV-signal, sig-
nal drift or instability
Check if the signal is still noisy with
water.
Impure buffer
Purge the pump according to
Pump User Manual. Run a system
wash with buffer.
Air in the pump or in
the UV cell
50 ÄKTAprime plus Operating Instructions 28-9597-89 AB
6 Troubleshooting

Related product manuals