What to do if my GE Laboratory Equipment pressure alarm is issued?
- DDaniel BennettJul 26, 2025
If the GE Laboratory Equipment issues a pressure alarm, it might be because the pressure is too high due to running the system at a flow rate that's too high for the column being used. Try lowering the flow rate. Alternatively, there may be a kink in the tubing, overtightened connections, or precipitation in the flow path. In this case, replace the tubing.