9.2 Chemical resistance specifications
Introduction
This section provides detailed information about chemical resistance of the ÄKTA go in-
strument to some common aggressive chemicals used in liquid chromatography. For
information regarding chemicals not listed in this section, contact your GE Healthcare
representative.
Refer to Safety Data Sheets (SDS) for information regarding characteristics,
human and environmental risks and preventive measures for chemicals used.
Make sure that you have the SDS available from your chemical distributor
and/or databases on the internet.
Note:
Scope
The information in this section applies to the ÄKTA go flow path. Recommended solutions
for the pump rinsing system are aqueous buffers or 20% ethanol. Do not use other
chemicals in the pump rinsing system.
Chemical resistance, long term
The following chemicals are suitable for continuous use.
EC noCAS noConcentrationChemical
N/AN/AN/AAqueous buffers, pH 2 to 12
200-662-267-64-110%Acetone
231-635-37664-41-730%Ammonia
235-186-412125-02-92 MAmmonium chloride
231-984-17783-20-23 MAmmonium sulfate
200-811-174-79-32 MArginine
202-859-9100-51-62%Benzyl alcohol
200-664-367-68-55%Dimethyl sulfoxide (DMSO)
229-998-86892-68-8100 mMDithioerythritol (DTE)
222-468-73483-12-3100 mMDithiothreitol (DTT)
200-837-375-08-120%Ethanol
ÄKTA go Operating Instructions 29360951 AC 103
9 Reference information
9.2 Chemical resistance specifications

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTA GO and is the answer not in the manual?

GE AKTA GO Specifications

General IconGeneral
BrandGE
ModelAKTA GO
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals