4.2.3 Prepare the pump rinsing system
Introduction
The pump rinsing system protects the pump seals from damage caused by precipitated
buffer remaining in the system. The seal prevents leakage between the pump chamber
and the drive mechanism of the pump.
Illustration of the pump rinsing
system
The illustration below shows the parts and tubing of the pump rinsing system.
1
2
3
4
DescriptionPart
Inlet tubing1
Outlet tubing2
Rinsing solution tube holder3
Rinsing solution tube4
48 ÄKTA go Operating Instructions 29360951 AC