Prime the pump rinsing system
Follow the instructions below to fill the pump rinsing system with rinsing solution.
ActionStep
Unscrew and remove the rinsing solution tube from the holder.1
Fill the rinsing solution tube with 50 mL of 20% ethanol or aqueous buffer.2
Screw the rinsing solution tube back into the holder.3
Insert the inlet tubing into the solution in the rinsing solution tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing liquid
tube.
4
Connect a 25 to 30 mL syringe to the outlet tubing. Draw liquid slowly into
the syringe until the rinsing system tubing is filled.
5
ÄKTA go Operating Instructions 29360951 AC 49