Questions and Answers

  • J
    John MeyerAug 15, 2025
    Why are there light leaks on my GE ImageQuant LAS 4000 Laboratory Equipment image?
    • G
      Gregory SmithAug 15, 2025
      If the ImageQuant LAS 4000 door is not completely closed, close it and expose the image again. Additionally, avoid direct sunlight to prevent light leaks on the image.
  • G
    Gregory MezaAug 18, 2025
    What to do if my ImageQuant LAS 4000 beeps and the Busy LED blinks?
    • C
      Carrie WaltersAug 18, 2025
      If the door of the ImageQuant LAS 4000 is open, close it. The subsequent processing should then start.
  • J
    Jennifer GomezAug 25, 2025
    Why is there unevenness seen on the image from my GE ImageQuant LAS 4000 Laboratory Equipment?
    • A
      Amy HamiltonAug 26, 2025
      If you see unevenness on the image, check that both left and right Epi light sources are correctly in place and that they are of the same type. If a lamp in the UV trans illuminator is broken, turn the power off and contact your GE Healthcare representative. Also, make sure that the Flat Frame is correctly exposed, and if not, expose it again.
  • H
    Heather DayAug 28, 2025
    What to do if my GE ImageQuant LAS 4000 Laboratory Equipment camera head makes a strange noise, emits smoke or generates a strange odor?
    • T
      tmaciasAug 28, 2025
      If the camera head generates a strange noise, emits smoke or generates a strange odor, turn off the ImageQuant LAS 4000 power switch immediately and contact your GE Healthcare representative due to a potential camera head failure.
  • N
    Nicole GutierrezAug 31, 2025
    What to do if my GE Laboratory Equipment camera head emits smoke?
    • M
      Michael NelsonAug 31, 2025
      Turn off the ImageQuant LAS 4000 power switch immediately and contact your GE Healthcare representative.
  • L
    Laura PetersSep 4, 2025
    What to do if my GE Laboratory Equipment camera head generates a strange odor?
    • J
      Jacob AustinSep 4, 2025
      Turn off the ImageQuant LAS 4000 power switch immediately and contact your GE Healthcare representative.
  • K
    Kevin MckenzieSep 7, 2025
    Why does the image appear misty on my GE Laboratory Equipment?
    • J
      Jennifer HoganSep 7, 2025
      If the image appears misty, it may be due to dew condensation in the optical system. Quit the ImageQuant LAS 4000 Control Software and wait until the operation environment meets the required specifications.
  • J
    John ThomasSep 12, 2025
    Why is there noise on an image exposed using UV trans tray on my GE Laboratory Equipment?
    • L
      Laura GentrySep 12, 2025
      If there is noise on an image exposed using a UV trans tray, it may be due to degradation of the UV trans tray. Purchase a new UV trans tray.
  • F
    fcoxNov 20, 2025
    How to fix if an appropriate image does not appear on my GE Laboratory Equipment?
    • K
      Kevin MolinaNov 21, 2025
      If an appropriate image does not appear, consider the following possibilities: * The image might be overexposed, so shorten the exposure time or check the iris setting. * The image might be underexposed, so lengthen the exposure time or check the iris setting. * The focus may not be correctly adjusted, so adjust the focus. * The tray, lens, or filter might be dirty, so clean them. * The wrong filter or lens might be installed, so install the correct ones. * The wrong sample tray might be used, so change to the correct sample tray. * The wrong light source might be selected, so select the correct one. * The Flat Frame file might not be exposed correctly, so expose it again. * The size of the object to be exposed might not coincide with the exposed area, so place the samp...

Summary

1 Introduction

1.1 Important user information

Covers critical safety instructions, intended use, and warnings for safe operation of the system.

2 The ImageQuant LAS 4000

3 Exchanging accessory parts

4 Operating the ImageQuant LAS 4000

4.1 Layout of the ImageQuant LAS 4000 Control Software

Shows the main window of the control software and identifies key features and controls.

4.2 Preparation for exposure

Outlines the steps for preparing the instrument and software before starting an exposure.

4.4 Placing the sample

Describes how to select the correct tray and position samples for optimal image acquisition.

4.13 Saving exposed images

Describes methods for saving captured images, including file formats and options.

5 Software reference guide

5.8 Method;Tray position

Guides users on selecting detection methods, light sources, filters, and tray positions.

5.12.1 CCD Calibration

Explains the process of creating correction files for CCD detector characteristics.

5.12.2 Flat Frame Calibration

Details how to create a flat frame correction file for optical characteristics.

6 Installing ImageQuant LAS 4000 Control Software

6.2 Install ImageQuant LAS 4000 Control Software under Windows XP

Step-by-step guide for installing the software on Windows XP systems.

6.3 Install ImageQuant LAS 4000 Control Software under Windows Vista

Step-by-step guide for installing the software on Windows Vista systems.