F1.8 24mm wide view lensF0.85 43mm LAS High Sens. lens
3.2.1 Removing the LAS High Sens. lens
1
Close the instrument door. If the power is turned off and the filter changer is down, proceed to step
4.
2
Select Option:Lens Exchange.
A message is displayed.
3
Click OK.
4
Turn off the instrument using the power switch.
ImageQuant LAS 4000 User Manual 28-9607-42 AB 19
3 Exchanging accessory parts
3.2 Installing or exchanging the lens
3.2.1 Removing the LAS High Sens. lens