4
Push the desired light source in along the guide at the bottom of the IDX.
5
Turn the lever at the front to lock the light source into position.
3.3.2 Changing the Epi lights
This section describes how to change a set of Epi lights. The RGB module Epi lights are also installed
as described here.
Note:
The tray should be at tray position 4 before installing Epi lights. If the tray is higher, then
it should be lowered to tray position 4 using the ImageQuant LAS 4000 Control Software
before attaching the lights, as described below.
Hook
Latch
Epi light base
1
If the tray is at Tray position 4, proceed to step 3. Otherwise, close the IDX door.
2
Click on Method/Tray position, set Tray position to 4, and click the OK button.
3
Open the IDX door. If no Epi light is currently in place inside the instrument, proceed to step 7.
ImageQuant LAS 4000 User Manual 28-9607-42 AB 29
3 Exchanging accessory parts
3.3 Changing the light sources
3.3.1 Changing the UV transilluminator or white light table

Table of Contents

Other manuals for GE ImageQuant LAS 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ImageQuant LAS 4000 and is the answer not in the manual?

GE ImageQuant LAS 4000 Specifications

General IconGeneral
BrandGE
ModelImageQuant LAS 4000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals