Manual: Select a suitable exposure time from the drop-down list. Alternatively select a time unit
and type in an exposure time.
When exposure type is increment or repetition
Choose a time unit from the drop-down list and enter an interval time.
TIP:
The interval time can be set from 10 seconds to 2 hours. For a UV transmitting light
source, the interval time can be set from 10 seconds to 1 minute.
Note:
An image can be exposed for up to 30 hours in Precision mode. However, the image
quality and quantitative performance are only guaranteed for a maximum of 2 hours.
5.3 Add digitization image
A chemiluminescence and a white-light epi-illuminated image are exposed by single-click operation
when this function is set to ON. After exposure, both chemiluminescence and white epi-illuminated
images can be saved.
This function can be used only when the method is chemiluminescence.
82 ImageQuant LAS 4000 User Manual 28-9607-42 AB
5 Software reference guide
5.2 Exposure time