Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-56
1. Remove the rear cover of the C-Arm. Release the three screws and open the package box. The
parts are on the back of the package.
2. Disconnect all cables connected to the power supply. Loose the screws and replace a new part.
3. Connect the cables according to the labels. Reassemble all parts.
4. Power on the system. Verify that the Workstation and C-Arm complete software boot and
initialization with no errors reported on their displays.
DC bus capacitor (P/N: 5329278)
1. Remove the rear cover of the C-Arm. Release the three screws and open the package box. The
part is on the back of the package.
There are two capacitors here but only one in FRU kit. Please identify the trouble for the part to determine
the quantity of FRU parts.
MISC, DC POWER
SUPPLY +5V,
+15V,-15V AND
+24V (P/N:
5085800)
MISC, DC POWER
SUPPLY +5V, +12V
AND +24V (P/N:
5085801)