Replacement
7-59
Power filter in Workstation
1. Remove the Workstation rear cover.
2. Unplug the cable connected to the filter.
3. Loose the screws and remove the filter.
Replace a new part and connect the cable
according to the schematic on the filter.
4. Reassemble the parts. Power on the system.
5. Verify that the Workstation and C-Arm
complete software boot and initialization with
no errors reported on their displays.
Button and Switch Kit (P/N: 5075922)
This kit includes 4 parts, Battery Power Switch (5085833), MISC, WKS POWER ON BUTTON (5085810), MISC,
C_ARM KEY SWITCH (5085817), Power on Indicator (P/N: 5085836)
Battery Power Switch (5085833):
1. Power off system, and unplug the power cord form
socket.
2. Remove the rear cover of the Workstation.
3. Unplug the wires connected to the Battery Power Switch
4. Release the secure ring of Battery Power Switch.
5. Remove Battery Power Switch and replace a new one.
6. Connect the AC cord to the socket.
7. Power on system, and then turn of the power on breaker,
press Battery Power Switch to verify UPS can be turned
off.
Fixed Ring
Connector