Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-14
MF_PWR
mainframe power detection
signal
PJ8-8 digital 0/+12V (>6V=High)
PDU_CONNECT_2
connection status signal
from PDU to SIB
PJ8-9 digital 0/+12V (>6V=High)
PDU_CONNECT_1
connection status signal
from PDU to SIB
PJ8-10 digital 0/+12V (>6V=High)
Invalid key beep command PJ8-19 digital 0/+12V (>6V=High)
INTLC_BOARD
interlock board connection
status signal
PJ8-20 digital 0/+12V (>6V=High)
ACK_PDU2SIB
acknowledge signal
between PDU and SIB
PJ8-21 digital 0/+12V (>6V=High)
SYS_PWR_OFF_CMD
PJ8-22 digital 0/+12V (>6V=High)
SYS_PWR_OFF_DENY
system power off deny
command
PJ8-23 digital 0/+12V (>6V=High)
MF_PWR_ON_CMD
mainframe power on
command
PJ8-24 digital 0/+12V (>6V=High)
UPS interface
UPS switch on/off control signal interface is shown as follows:
UPS battery/line status signal
UPS switch off interface in emergency is shown as follows:
UPS EMERGENCY SWICH
OFF BUTTON
PJ15-1 Digital 3.3V/5V TTL
PJ15-2 Digital 3.3V/5V TTL
Test Point: