Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-28
button has pressed
HAND_HLF_SECURITY
Hand security line
termination
Passive
Open/Close
J1-C27
-
HAND_HLF
High Level Fluoro
hand switch
termination
Passive
Open/Close
J1-B7
HAND_HLF & HAND_COM is
connected means HAND_HLF
button has been pressed
HAND
_HLU_SECURITY
High Level Fluoro
hand security line
termination
Passive
Open/Close
J1-A22
-
HAND_COM
Common
termination
Passive
-
J1-C15
-
Foot Switch
Signals from foot switch to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
FOOT_FLU
Foot switch
termination
Passive
Open/Close
J1-A27
FOOT_FLU & FOOT_COM is
connected means FOOT_FLU
button has been pressed
FOOT_FLU_SECURIT
Y
Foot security line
termination
Passive
Open/Close
J1-C28
-
FOOT_ HLF
High Level Fluoro
foot switch
termination
Passive
Open/Close
J1-A14
FOOT_HLF & FOOT_COM is
connected means FOOT_HLF
button has been pressed
FOOT_
HLF_SECURITY
High Level Fluoro
foot security line
termination
Passive
Open/Close
J1-C31
-
FOOT_BUTTON1
Button 1
Passive
Open/Close
J1-A20
FOOT_BUTTON1 & FOOT_COM is
connected means FOOT BUTTON 1
has been pressed
FOOT_BUTTON2
Button 2
Passive
Open/Close
J1-A19
FOOT_BUTTON2 & FOOT_COM is
connected means FOOT BUTTON 2
has been pressed
FOOT_COM
Common
termination
Passive
-
J1-B32
-
Emergency Switch
Signals from emergency switch to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
EMERGENCY_SWT1
One termination of
Emergency
Passive
Open/Close
J1-C11
EMERGENCY_SWT1 &
REF_EMERGENCY_SWT1 is not

Table of Contents