EasyManua.ls Logo

GE Brivo OEC 865 - Imaging Modes

GE Brivo OEC 865
551 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Overview
2-29
switch1
connected means
EMERGENCY_SWT has been
pressed
REF_
EMERGENCY_SWT1
The other
termination of
Emergency
switch1
Passive
-
J1-B11
-
EMERGENCY_SWT2
One termination of
Emergency
switch2
Passive
Open/Close
J1-C25
EMERGENCY_SWT2 &
REF_EMERGENCY_SWT2 is not
connected means
EMERGENCY_SWT2 has been
pressed
REF_
EMERGENCY_SWT2
The other
termination of
Emergency
switch2
Passive
-
J1-A26
-
Removable Filter
Signals from removable filter to MCB are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
PROTECT_A
Removable filter A
sensor signal
Digital
0V/+15V
J1-B31
+15V= Filter removed
PROTECT_B
Removable filter B
sensor signal
Digital
0V/+15V
J1-B30
+15V= Filter removed
Indicator
Signals from MCB to indicator are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
XRAY_INDICATOR_A
Anode of indicator
Passive
-
J1-B8
The indicator will be lighted when a
certain current about 12mA flows
from XRAY_INDICATOR_A through
indicator to XRAY_INDICATOR_B.
XRAY_INDICATOR_B
Cathode of indicator
Passive
-
J1-A8
Program Port
Program port signals are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
SERIAL_TXD
Transmit signal of serial port
Digital
-7V-+7V
J1-A31
-
SERIAL_RXD
Receive signal of serial port
Digital
-7V-+7V
J1-C22
-
Interlock Loop
Interlock loop signals are listed as follows.
Name
Description
Attribution
Scale
Pin
Comment
INTERLOCK_LOOP_A
One termination of
interlock loop
Passive
-
J2-B16
INTERLOCK_LOOP_A &
INTERLOCK_LOPP_B is connected in

Table of Contents