2048723-002C CARESCAPE V100 Vital Signs Monitor 3-13
Installation: Configuring the monitor
NOTE
The first tone sounds as you first press the Menu button and the second
tone sounds after the monitor has saved the settings.
After all options pages have been displayed, the display will return to the first
service mode page (initial calibration page). The Service Mode option pages are
as follows:
Displayed on monitor Function
0 (in min window) Refer to calibration section for functions.
1 (in min window) Refer to calibration section for functions.
2 (in min window) Refer to calibration section for functions.
3 (in min window) NIBP Algorithm Type loaded
(Displayed in MAP/Cuff display window)
1: DINAMAP Classic NIBP
2: DINAMAP Auscultatory NIBP
3: DINAMAP SuperSTAT NIBP
NOTE
Changing setting effects NIBP performance.
3 (in min window) Alarm mode
(Displayed in Pulse Rate window)
0: IEC mode
1: Legacy mode
NOTE
Use the Alarms button to toggle between
the available options.
Refer to the “Alarms” section of the
operators manual for a detailed
description of the alarm modes.
4 (in min window) SpO
2
Type loaded
(Displayed in SpO2% display window)
0: No SpO
2
1: Nellcor
2: Masimo
3: Ohmeda
NOTE
Incorrect setting will cause fatal 930 alarm
during operation.