3-14 CARESCAPE V100 Vital Signs Monitor 2048723-002C
Installation: Configuring the monitor
NOTE
You must be in option page 6 (in min window) to save any changes made in
service mode.
5 (in min window) Temperature loaded
(Displayed in Temperature display window)
0: No Temp
1: Alaris Turbo Temp
2: Alaris Tri-Site
3: Exergen
6 (in min window) Language
The number displayed in Pulse Rate display
window indicates the language setting. These
range from 0 to 20. For example, 0 indicates
English. The language setting is used in printed
reports. Russian, Greek, Korean, and Japanese are
printed in English only.
0 English
1 Not used
2 Czech
3 Danish
4 Dutch
5 Finnish
6 French
7 German
8 Greek
9 Hungarian
10 Italian
11 Japanese
12 Korean
13 Norwegian
14 Polish
15 Not used
16 Portuguese
17 Russian
18 Slovak
19 Spanish
20 Swedish