5-26 120 Series Maternal/Fetal Monitor Revision B
2015590-001
Setup Procedures: Service Mode Screens
Install Options Screen
Use the Install Options screen (Figure 5-15) to:
select the default settings used when the 120 Series Monitor is turned on;
select the line frequency;
enable/disable ECG artifact elimination;
configure the recorder’s vertical scale;
select the language used on the display and printed on the strip chart paper;
enable/disable a one-minute blood pressure interval selection on the maternal
BP Setup screen;
enable/disable the secondary heart rate offset feature;
select the uterine activity baseline value used when a tocotransducer is in use
and you press the
UA Reference pushbutton;
set the type of mark printed when a patient presses the button on the Remote
Event Marker accessory;
enable/disable Corolan address checking circuitry;
enable/disable the heartbeat coincidence feature;
enable/disable the Smart BP feature;
select between real-time and chart-style vital signs history intervals;
select the recorder font size;
select between fetal alerts and alarms (if the Spectra Alerts option is installed);
enable/disable the alert suspend feature (if the Spectra Alerts option is
installed);
print all the setup settings on the strip chart recorder;
print the error log (refer to “Chapter 8, Self-Tests”);
access the Communications Setup screen; and
access the Diagnostic Control screen.
Figure 5-15. Install Options Screen
INSTALL OPTIONS
DEFAULT SETTINGS: FACTORY
LINE FREQUENCY: 60 HZ
ECG ARTIFACT ELIMINATION: OFF
SCALING: 30-240
LANGUAGE: ENGLISH
NBP 1 MIN INTERVAL: OFF
HR OFFSET: 10 MIN
DEFAULT TOCO REFERENCE: 10
FM REMOTE MARK: ON
COROLAN ADDRESS CHECKING: OFF
HBC: OFF SMART BP: OFF
VS PRINT INTERVAL: REAL TIME
RECORDER FONT SIZE: SMALL
FETAL ALERT/ALARM: OFF
ALERT SUSPEND: OFF
PRINTALL LOG COMM
TESTS EXIT