EasyManua.ls Logo

GE CRxFlex - Environmental Conditions; Electrical Data

GE CRxFlex
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CRxFlex Operating Manual V1.1 37
Environmental Conditions
Operation
Temperature recommended:
20 °C - 25 °C
Temperature allowed
15 °C - 30 °C
Maximum temperature change
0.5 °C/min.
Relative humidity
recommended
30 % - 60 %
Relative humidity allowed
15 % - 75 % (non condensed
Magnetic field
compliant with EN 61000-4-8,
Level 2
Sunlight exposure
Not to be operated in direct
sunlight, max. 2000 lux
Transport
IEC60721-3-2 (1997: class 2K2
and 2M2, with following
restrictions
Temperature
-25 °C to +55 °C
Vibration
5 200 Hz (vertical, longitudinal,
transversal axis
Electrical Data
Operating voltage
Auto-ranging power supply from
100-120 / 220-240 V
Operating current
2.0 / 1.0 A
Mains frequency
50 / 60 Hz
Mains fuse protection
Europe: min. 10 A, max. 16 A
USA & Japan: min. 10 A, max.
15A
Power consumption
Standby
max. 60W
During operation
max. 220W
Dimensions in cm [inch]
78.6 [30.9] 69.3 [27.2]
52.5 [20.7]
Dimensions in cm [inch]
69.3 [27.2]
52.5 [20.7]
100.2 [39.5]

Related product manuals