EasyManua.ls Logo

GE DigitalFlow DF868

GE DigitalFlow DF868
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1. Calibration
12 DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.7.3 Entering Pulse Number
5. Enter the number of pulses desired (between 1 and 10,000) and press the [ENT] key. That number of pulses will
then be output at the specified frequency.
6. Repeat steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the test,
contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
To calibrate/test additional slots, press [F4] (EXIT) and proceed to the appropriate section.
To leave the Calibration Menu, press [EXIT] twice.

Table of Contents

Other manuals for GE DigitalFlow DF868

Related product manuals