13
LOSS OF PULSE: The monitor will detect loss of pulse from patient and
enter a no signal state within 10 seconds.
Mechanical
DIMENSIONS: Height: 9.1 in (23.0 cm)
Width: 7.3 in (18.5 cm)
Depth: 6.9 in (17.5 cm)
WEIGHT including battery: 8.3 lb (3.75 kg)
MOUNTINGS: Self-supporting on rubber feet or pole mountable.
PORTABILITY: Carried by recessed handle or pole mounted.
CLASSIFICATION INFORMATION: • Mode of Operation - Continuous
• Degree of Protection against harmful ingress of water: (See
IPX1 definition).
Power Requirements:
POWER CONVERTER: Protection against electrical shock - Class 1
INPUT VOLTAGE: 115 / 230 VAC, 50 / 60 Hz (nominal),
90 ~ 264 VAC, 47 ~ 63 Hz (range)
MONITOR: Protection against electrical shock - Class 1
INPUT VOLTAGE: 24 VDC (nominal), 12-30 VDC from supplied power converter
EXTERNAL DC LINE FUSE: T3.15A 250 VAC
BATTERY: 12 volt, 2.3 amp-hours.
Minimum operation time: 2 hours (5 minute auto cycle with adult
cuff at 25°C with power save mode enabled) from full charge.
Time for full recharge: 8 hours from full discharge
Environmental
OPERATING TEMPERATURE: + 5° C to + 40° C (+ 41° F to + 104° F)
OPERATING ATMOSPHERIC
PRESSURE RANGE: 700 to 1060 hectoPascal
STORAGE TEMPERATURE: –20° C to + 50° C (– 4° F to + 122° F)
STORAGE / TRANSPORTATION
ATMOSPHERIC PRESSURE: 500 to 1060 hectoPascal
HUMIDITY RANGE: 0 % to 95 % non-condensing
RADIO FREQUENCY Complies with IEC Publication 801-3 (Draft 7, second edition)
ELECTROMAGNETIC IMMUNITY: 3.0V/m, modulation 80% AM @ 1kHz, frequency range
26.0~1000.0Mhz