Druck ADTS 405 Service Manual
K171 Issue No. 1 Page 1-4
SPECIFICATION
Electrical power supply ......................................................................................................................................... 88 to 260V a.c.
........................................................................................................................................................ 47 to 440Hz
ADTS 405 ........................................................................................................................ 200 VA maximum
ADTS 405F........................................................................................................................500VA maximum
Pneumatic outputs
Altitude range .........................................................................................................................................-2000ft to +80000ft
Altitude accuracy
Sea level ................................................................................................................................................................... ± 5ft
30,000ft ....... ............................................................................................................................................................ ± 7ft
60,000ft .................................................................................................................................................................. ± 25ft
Static
range ...................................................................................................................................... 1355 to 35 mbar abs
accuracy....................................................................................................................................... ± 0,012% reading
............................................................................................................................................................ 0.008% FS
Airspeed ranges
civil ............................................................................................................................................ 0 to 500 knots
military .................................................................................................................................... 0 to 850 knots
special................................................................................................................................... 0 to 1000 knots
Airspeed accuracy
50 knots ...................................................................................................................................................... ± 0,5 knots
500 knots ................................................................................................................................................. ± 0,07 knots
Qc
range ...................................................................................................................................... 35 to 2700 mbar abs
accuracy....................................................................................................................................... ± 0,015% reading
............................................................................................................................................................ 0.007% FS
Special
range ...................................................................................................................................... 35 to 3500 mbar abs
accuracy....................................................................................................................................... ± 0,015% reading
............................................................................................................................................................ 0.007% FS
Mach number
civil ........................................................................................................................................................ 0 to 1,00
military ................................................................................................................................................ 0 to 6,00
Accuracy ....................................................................................................................................... Better than 0,005
Rate of climb
Generated ranges
civil ........................................................................................................................................ 0 to 6000 ft/min
military ............................................................................................................................. 0 to 15000 ft/min
Accuracy ................................................................................................................................................ ±1% of value
Dimensions
ADTS 405 ............................................................................................................. 485 mm x 2
70 mm x 305 mm
Weight .................................................................................................................................................................... 12 kg
ADTS 405F .......................................................................................................... 442 mm x 755 mm x 332 mm
Weight .................................................................................................................................................................... 29 kg
Temperature
Operating ............................................................................................................................................... 10° to +50°C
Calibrated ................................................................................................................................................. 5° to +35°C
Storage.................................................................................................................................................. -40° to +70°C
Note: The rates, ranges and limits may be different depending on customer requirements.