Druck ADTS 405 Service Manual
K171 Issue No. 1 Page 5-5
Completion
18. When the calibration check is complete, repeatedly use the QUIT key to return to the first
calibration screen.
i. If necessary, carry out a rate calibration or press QUIT can be pressed again to exit
calibration mode.
ii. Select go to ground or stay at the current pressure.
Point
Calibration Pressure
inHg mbar
1 1 34
2 5.1 173
3 10.2 345
4 20.4 691
5 30.5 1033
6* 38.7 1311
7« 50.9 1724
8« 63.1 2137
9« 77.4 2621
10« 89.6 3034
11« 103.4 3502
Table 5-1 Table 5-2
Calibration Check Points Ps Calibration Check Points
(both channels)
Point
Calibration Pressure
inHg mbar
1 1 34
2 5.1 173
3 10.2 345
4 15.3 518
5 20.4 691
6 25.4 860
7 30.5 1033
8* 38.7 1311
9« 50.9 1724
10« 63.1 2137
11« 80 2709
Point
Calibration Pressure
inHg mbar
1 1 34
2 3.1 105
3 5.1 173
4 7.1 240
5 10.2 345
6 15.3 518
7 20.4 691
8 25.4 860
9 30.5 1033
10 34.6 1172
11 40 1355
Table 5-3
Calibration Check Points
Special version
(both channels)
* Ps END
« Pt value only