GE LOGIQ V5/LOGIQ V3
D
IRECTION 5496012-100, REVISION 3 BASIC SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 21
CF/PDI Auto
Sample Volume
Yes Set the default value at Utility -> Imaging -> CF Mode.
CF/PDI Focus
Depth
Yes
CF/PDI Frequency Yes
CF/PDI Auto
Frequency
Yes
CF/PDI Center
Depth
Yes
PDI Yes Power Doppler Imaging (PDI) is a color flow mapping
technique used to map the strength of the Doppler signal
coming from the tissue rather than the frequency shift of
the signal. Using this technique, the ultrasound system
plots color flow based on the number of reflectors that are
moving, regardless of their velocity. PDI does not map
velocity, therefore it is not subject to aliasing.
TVI (Option) Yes Tissue Velocity Imaging (TVI) calculates and color-codes
the velocities in tissue. The tissue velocity information is
acquired by sampling of tissue Doppler velocity values at
discrete points.The information is stored in a combined
format with gray scale imaging during one or several
cardiac cycles with high temporal resolution.
TVD Yes TVD: Tissue Velocity Doppler: basing on TVI mode,
activate a sample volume of PW ventricular wall to get the
spectral information of the sample section.
Table 4-5 Color Flow Mode Controls
Control Possible Bioeffect Description/Benefit

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals