Lullaby Warmer
Maintenance and Service Manual -English
2042376-001
Rev G
GE Healthcare
LullabyWarmer
Maintenance and Service manual

Table of Contents

Questions and Answers

  • S
    Suzanne GibbsAug 25, 2025
    Why does the global alarm indicator continue to be lit up on GE Medical Equipment?
    • R
      russellpatrickAug 25, 2025
      If your GE Medical Equipment displays a global alarm that continues to be lit, it is likely due to a CPU failure. To resolve this, replace the PWA with FRU 2043951-001 for warmers with a halogen lamp and transformer or replace the PWA with FRU 2062277-001 for warmers with SMPS and LED observation lamps.
  • S
    shannonbridgesAug 28, 2025
    What should I do if the skin probe fails on my GE Medical Equipment?
    • T
      Theresa EdwardsAug 28, 2025
      If your GE Medical Equipment's skin probe fails, it could be due to a defective probe, so replace the skin probe. Alternatively, the PWA might be defective; replace the PWA with FRU 2043951-001 for warmers with a halogen lamp and transformer, or replace the PWA with FRU 2062277-001 for warmers with SMPS and LED observation lamp.
  • W
    William AndersonAug 30, 2025
    What does it mean when 'key' is displayed on GE Medical Equipment LullabyTM Warmer?
    • D
      Derek GarciaAug 30, 2025
      If the LullabyTM Warmer displays "key" on the Control Temperature display with alarm indication when switched on, it's likely because a key on the front panel was pressed during startup. To resolve this, avoid pressing any key when switching on the system. Also, check if any of the keys on the front panel are stuck and ensure they are not jammed.
  • K
    Kyle HughesAug 31, 2025
    Why does the heater hood on my GE Medical Equipment not swivel?
    • E
      Erica AndersonSep 1, 2025
      If the heater hood of your GE Medical Equipment does not swivel, it might be because the hood is not aligned to the center. To fix this, align the heater hood to the center and pull the swivel knob.
  • R
    Ruben HensleySep 3, 2025
    What does 'CFg' mean on the GE Medical Equipment Control Temperature display?
    • A
      Alyssa LivingstonSep 3, 2025
      If your GE Medical Equipment displays "CFg" on the Control Temperature display, it indicates that the configuration switch is configured wrongly. To resolve this, ensure that the configuration switch SW1 on the PWA is configured as per the table.
  • S
    Stephen DouglasSep 5, 2025
    What does 'rLy' mean on the GE Medical Equipment LullabyTM Warmer display?
    • W
      warrenjohnSep 5, 2025
      If the LullabyTM Warmer displays "rLy" on the Control Temperature display with alarm indication when switched on, it indicates that the safety relay is defective. To resolve this, replace the PWA with FRU 2043951-001 for warmers with a halogen lamp and transformer, or replace the PWA with FRU 2062277-001 for warmers with SMPS and LED observation lamp.
  • D
    Diana PerrySep 7, 2025
    Why does my GE Medical Equipment have no mobility?
    • M
      moniquewilsonSep 7, 2025
      If your GE Medical Equipment has no mobility, check if the castor brake is locked and release it. If the castor is damaged, replace it.
  • S
    Shaun JohnsonSep 8, 2025
    What to do if the base leg cap is loose on GE Medical Equipment?
    • S
      Susan GomezSep 9, 2025
      If the base leg cap of your GE Medical Equipment is loose, it may be due to a broken leg cap or an inability to tighten the cap; in this case, replace the leg cap. Alternatively, a loose screw could be the cause; align the cap and tighten the screw.
  • G
    grossbryceSep 12, 2025
    What does 'trC' mean on the GE Medical Equipment LullabyTM Warmer display?
    • A
      Andrea MartinezSep 12, 2025
      If the LullabyTM Warmer displays "trC" on the Control Temperature display with alarm indication when switched on, it means that the heat control triace is defective. To fix this, replace the PWA with FRU 2043951-001 for warmers with a halogen lamp and transformer, or replace the PWA with FRU 2062277-001 for warmers with SMPS and LED observation lamp.
  • S
    Suzanne RoseSep 13, 2025
    What causes the push button key failure in GE Medical Equipment?
    • K
      Katherine MartinSep 13, 2025
      If the push button key on your GE Medical Equipment fails, the cause is likely a defective push button key. To resolve this, replace the PWA.

Summary

Important Safety Information

Warnings, Cautions, and Notes

Introduces general warnings, cautions, and notes with symbols for hazard awareness.

About this Manual

Scope and Intended Users

Details the manual's purpose, target audience, and the warmer's intended applications.

Organization of this Manual

Explains the structure of the manual and the content of each chapter.

Chapter 1 Introduction

Indications for Use

Describes the intended use and applications of the warmer for neonatal thermoregulation.

Chapter 2 System Setup and Checkout Procedures

Mechanical Checks

Details procedures for checking the warmer's mechanical components and connections.

Controller Checks

Outlines checks for the control panel functions, displays, and switches.

Accessories Check

Describes how to check the proper mounting and functionality of accessories.

Chapter 3 Maintenance and Cleaning

Repair Policy

Defines procedures for warranty and non-warranty repairs, including contact information.

Required Tools and Personnel Protection Equipment

Lists necessary tools and PPE for maintenance and servicing.

Maintenance Schedule

Outlines recommended intervals for operator and service maintenance.

Cleaning Instructions

Provides guidelines for cleaning the warmer and its components, with notes on solutions.

Chapter 5 System Installation Procedure

Required Tools and Personnel Protection Equipment (PPE)

Lists necessary tools and PPE for system installation.

Contents of LullabyTM Warmer Packaging

Details the items included in the warmer's packaging.

Unpacking the Shipping Box

Provides instructions for unpacking the warmer from its shipping box.

Pillar to Leg Assembly and Height Adjustment

Guides the assembly of the pillar to the leg and height adjustment.

Assembly of Pillar and Bed

Details the process of assembling the pillar and the bed components.

Assembling the Pillar and the Hood

Explains how to assemble the pillar and the hood sections.

Connecting the Power Cord

Instructs on how to connect the warmer to the power source.

Mounting the Accessories

Guides the installation of various optional accessories.

Chapter 6 System Dismantling Procedure

Dismantling the Accessories

Provides instructions for removing accessories from the warmer.

Dismantling the Base and Bed

Details the procedure to dismantle the base and bed components.

Dismantling the Hood Assembly

Guides the removal of the hood assembly from the unit.

Dismantling the Pillar to Leg Assembly

Explains how to dismantle the pillar from the leg assembly.

Chapter 7 Field Replaceable Unit (FRU) Replacement Procedures

Replacing the Parts of the LullabyTM Warmer

General overview of replacing FRUs in the warmer.

Chapter 9 Illustrated Parts

Hood Assembly

Illustrates the components of the hood assembly.

Base Assembly

Shows exploded views of the base assembly components.

Bed Assembly

Illustrates the components of the bed assembly.

Appendix B Specifications

Power Requirements

Details the electrical power requirements of the warmer.

Standards

Lists the safety and performance standards the device meets.

Classification

Describes the equipment classification regarding safety and ingress protection.

Operating Environment

Specifies the recommended environmental conditions for operation.

Storage and Transport Conditions

Outlines conditions for storing and transporting the device.

User Control Settings

Describes the ranges for user-adjustable settings.

Performance

Details accuracy and performance characteristics of the warmer.

Mechanical Specifications

Lists the physical dimensions and weight of the warmer.

Weight Bearing Capacity

Specifies the maximum load capacity for various parts.

Appendix C Electromagnetic Compatibility

Electromagnetic Emissions

Details RF emissions and compliance for the device.

Immunity Test

Lists immunity test levels and compliance criteria.

Recommended Separation Distances

Provides guidelines for separating RF equipment from the warmer.

Appendix E Assembling and Dismantling the MSB Drawer Assembly

MSB Drawer Assembly with Base Leg Cover

Instructions for assembling and dismantling the MSB drawer with base leg cover.

MSB Drawer Assembly with Drawer Clamp

Instructions for assembling and dismantling the MSB drawer with drawer clamp.

GE Lullaby Warmer Specifications

General IconGeneral
BrandGE
ModelLullaby Warmer
CategoryMedical Equipment
LanguageEnglish

Related product manuals