© 2010 by General Electric Company. All rights reserved. 2049809-001 Rev B 9
Chapter 2: Safety
authorized under the certication procedure in Part 2 of the rules. One or more frequency coordinators maintain
a database of all equipment used in conjunction with WMTS.
For more information visit http://www.fcc.gov
Industry Canada Information
The Mini Telemetry System complies with Industry Canada RSS-210 .
Transmitter Antenna
The Mini Telemetry Transmitter has been designed to operate with an antenna having a maximum gain of 6 dBi.
Antenna having a higher gain is strictly prohibited as per regulations of Industry Canada. The required antenna
impedance is 50 Ω.
Interference
Mini-telemetry System operation is subject to the following two conditions:
1. This device may not cause interference.
2. This device must accept any interference, including interference that may cause undesired operation of the
device.
Wireless Medical Telemetry Service
The Mini Telemetry System is only permitted for installation in hospitals and health care facilities. Devices shall
not be operated in mobile vehicles (even ambulances and other vehicles associated with health care facilities).
The installer/user of this device shall ensure that it is at least 80 km from the Penticton radio astronomy station
(British Columbia latitude: 49° 19’ 12” N: longitude: 118° 59’ 56” W). For medical telemetry systems not meeting
this 80 km separation (e.g. the Okinagan valley, British Columbia) the installer/user must coordinate with and
obtain the written concurrence of the Director of the Penticton radio astronomy station before the equipment
can be installed or operated. The Penticton contact is Tel: 250-493-2277/fax: 250-493-7767.
CE Marking Information Compliance
The Mini Telemetry System bears CE mark CE-0459 indicating conformity with the provisions of the Council
Directive 93/42/EEC concerning medical devices and fullls the essential requirements of Annex I of this directive.
The product is radio-interference protection class A in accordance with EN 55011
The country of manufacture can be found on the equipment labeling.