© 2010 by General Electric Company. All rights reserved. 2049809-001 Rev B 45
Appendix B
Transmitter
Category Technical Specications
ECG Mode
Input Impedance:
DC Tolerance:
Common Mode Rejection Ratio:
FECG Sensitivity:
MECG Sensitivity:
Leakage Current:
>1 GΩ
±1 V
>90 dB
30 µ V to 1 mV
0.5 mV to 5 mV
< 10 µA (TYPE CF)
TOCO Mode
Type:
Sensitivity:
Range:
Tocotransducer
20 µ V/relative unit
0 to +100 relative units
IUPC Mode
Type:
Sensitivity
Range:
Leakage Current:
DC Strain Gauge
20 µ V/mmHg
0 to +100 mmHg
< 10 µA (TYPE CF)
Antenna Type: Internal , helical
Control:
Power Key
Mark key
Speaker volume key (internal speaker and head set)