What to do if GE Measuring Instruments shows a flashing Absorbance reading of 2.00 Abs?
- MMatthew SnowAug 18, 2025
If you observe a flashing Absorbance reading of 2.00 Abs, it indicates that the sample needs to be diluted.

What to do if GE Measuring Instruments shows a flashing Absorbance reading of 2.00 Abs?
If you observe a flashing Absorbance reading of 2.00 Abs, it indicates that the sample needs to be diluted.
What to do if GE NanoVue Plus shows Error code 001 symbol?
If Error code 001 appears, it means the lamp needs changing.
Why does GE NanoVue Plus switch off after calibration?
The instrument may be switching off after calibration because the ON/OFF button is being held for too long. This can cause the instrument to receive both ON and OFF signals. Try adjusting the timing of your finger press at switch on.
How to fix GE Measuring Instruments failing start up calibration check?
If the instrument fails the start up calibration check, ensure that the sample and reference beams are clear when the instrument is switched on.
What to do if GE NanoVue Plus Measuring Instruments shows X against Lamp on initialisation screen?
If an 'X' appears against the lamp on the initialization screen, it indicates that the lamp needs to be replaced.
What to do if GE Measuring Instruments shows Absorbance readings unstable?
If you are experiencing unstable absorbance readings, ensure that Reference QA is switched on, background correction is enabled, and that a 2 µl sample is being used.
Why does GE NanoVue Plus show a negative reading?
Negative readings can occur if the Reference and Test cuvettes are mixed up.
What to do if GE NanoVue Plus instrument will not zero?
The instrument may not zero if the cuvette is placed in the wrong orientation or if the wrong cuvette material is used for the wavelength.
What to do if GE NanoVue Plus Measuring Instruments shows REF is displayed when T is pressed?
If 'REF' is displayed when 'T' is pressed, it means the instrument was not zeroed before a sample measurement was attempted.
What to do if GE Measuring Instruments gives Poor reproducibility on DNA concentration and purity measurements?
Poor reproducibility in DNA concentration and purity measurements can occur if the wrong cuvette is used (not a 15 mm beam height or wrong material not letting UV light pass through), if background correction mode is not switched on, or if the measured absorbance is too low (should be >0.1 A for accurate results).
| Brand | GE |
|---|---|
| Model | NanoVue Plus |
| Category | Measuring Instruments |
| Language | English |
This booklet covers the ranges of instruments including Ultrospec, Novaspec, GeneQuant, SimpliNano, and NanoVue.
Spectrophotometers measure sample absorbance using light. Applications include DNA/RNA concentration and purity.
Defines WARNING and CAUTION symbols indicating hazardous situations and potential injury.
Details safety directives (LVD, EMC, WEEE, ROHS, Machinery) and conformity standards for the equipment.
Highlights potential hazards including UV sources, high voltages, PC connections, and hot surfaces on the instruments.
Covers instrument inspection, environmental conditions, power requirements, fuse types, placement, ventilation, and on/off switch locations.
Details the initial steps for operating the instrument, including switching it on and selecting the appropriate software module.
Provides instructions on using cuvettes for sample measurement, including selecting the correct cuvette and zeroing the instrument.
Guides on performing measurements with micro volume instruments like SimpliNano and NanoVue Plus using sample ports or plates.
Outlines post-run procedures including cleaning cuvettes, sample ports, plates, and guidelines for sample disposal.
Details user maintenance tasks for various instrument models, such as lamp changes and cell holder cleaning.
Provides step-by-step instructions for changing the lamp on Novaspec III/Plus instruments.
Details the procedure for changing the lamp on Ultrospec 8000/9000 instruments, including safety precautions.
Provides guidance on emergency procedures in case of malfunction or hazard, including power disconnection and decontamination.
Outlines the requirement for decontamination of equipment before decommissioning and following local regulations.
States that spectrophotometers contain hazardous substances and directs users to local GE representatives for information.
Instructs on proper disposal of electrical and electronic equipment, advising separate collection and contact with authorized representatives.