EasyManua.ls Logo

GE Novaspec Plus - User Manual

GE Novaspec Plus
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Spectrophotometer Health &
Safety Document including
General Operating Instructions
Original Instructions

Questions and Answers

  • J
    James VargasAug 18, 2025
    What to do if GE Novaspec Plus Measuring Instruments shows a flashing Absorbance reading of 2.00 Abs?
    • T
      Theresa LawsonAug 18, 2025
      If you observe a flashing Absorbance reading of 2.00 Abs, it indicates that the sample needs to be diluted.
  • L
    lozanowayneAug 20, 2025
    What to do if GE Novaspec Plus Measuring Instruments shows Error code 001 symbol?
    • A
      Ann CooleyAug 20, 2025
      If Error code 001 appears, it means the lamp needs changing.
  • E
    Elizabeth SnyderAug 23, 2025
    Why does GE Measuring Instruments switch off after calibration?
    • W
      Wanda SmithAug 23, 2025
      The instrument may be switching off after calibration because the ON/OFF button is being held for too long. This can cause the instrument to receive both ON and OFF signals. Try adjusting the timing of your finger press at switch on.
  • K
    Katie AndersonAug 26, 2025
    How to fix GE Novaspec Plus Measuring Instruments failing start up calibration check?
    • S
      Stephen ElliottAug 26, 2025
      If the instrument fails the start up calibration check, ensure that the sample and reference beams are clear when the instrument is switched on.
  • A
    Angela JenkinsAug 27, 2025
    What to do if GE Novaspec Plus shows X against Lamp on initialisation screen?
    • M
      Matthew StephensAug 27, 2025
      If an 'X' appears against the lamp on the initialization screen, it indicates that the lamp needs to be replaced.
  • R
    Robert DavenportAug 30, 2025
    What to do if GE Novaspec Plus shows Absorbance readings unstable?
    • S
      smithkathrynAug 30, 2025
      If you are experiencing unstable absorbance readings, ensure that Reference QA is switched on, background correction is enabled, and that a 2 µl sample is being used.
  • R
    Robert KingSep 1, 2025
    Why does GE Novaspec Plus show a negative reading?
    • D
      Deborah NashSep 1, 2025
      Negative readings can occur if the Reference and Test cuvettes are mixed up.
  • G
    Gabrielle JonesSep 3, 2025
    What to do if GE Measuring Instruments instrument will not zero?
    • E
      Eugene DuranSep 3, 2025
      The instrument may not zero if the cuvette is placed in the wrong orientation or if the wrong cuvette material is used for the wavelength.
  • A
    Andrew ShermanSep 6, 2025
    What to do if GE Measuring Instruments shows REF is displayed when T is pressed?
    • T
      Tina HicksSep 6, 2025
      If 'REF' is displayed when 'T' is pressed, it means the instrument was not zeroed before a sample measurement was attempted.
  • J
    joseph49Sep 8, 2025
    What to do if GE Measuring Instruments gives Poor reproducibility on DNA concentration and purity measurements?
    • J
      joneserinSep 8, 2025
      Poor reproducibility in DNA concentration and purity measurements can occur if the wrong cuvette is used (not a 15 mm beam height or wrong material not letting UV light pass through), if background correction mode is not switched on, or if the measured absorbance is too low (should be >0.1 A for accurate results).

Summary

HEALTH AND SAFETY

Scope

Details the range of spectrophotometer instruments covered by this document.

Intended use

Explains the function of spectrophotometers and their typical applications in labs.

Safety Notices

Defines Warning and Caution symbols to alert users to potential hazards.

General Safety

Lists relevant EU directives and international standards the equipment complies with.

General Hazards

Highlights potential dangers including UV sources, high voltage, and hot surfaces.

Unpacking and installation

Provides initial steps for inspecting and setting up the instrument.

OPERATING AND MAINTENANCE

Preparation before starting

Outlines the procedure for powering on the instrument and selecting software modules.

Performing a run - cuvettes

Details the process of using cuvettes for sample measurement.

Performing a run - micro volume instruments

Explains sample measurement using micro volume ports on specific models.

Post run procedures

Describes cleaning steps for cuvettes, sample ports, and plates after use.

User Maintenance

Lists maintenance tasks that users can perform for different instrument types.

Novaspec III;Plus Lamp Change

Step-by-step instructions for replacing the lamp in Novaspec III/Plus models.

Ultrospec 8000;9000 Lamp Change

Step-by-step instructions for replacing the lamp in Ultrospec 8000/9000 models.

TROUBLESHOOTING

Emergency Procedures

Actions to take in case of a malfunction or spill of hazardous material.

Recycling Procedures

Guidelines for decontaminating and scrapping equipment according to regulations.

Recycling of hazardous substances

Information regarding hazardous substances within the spectrophotometers.

Disposal of electrical components

Proper methods for disposing of electronic waste.

Specifications

Technical specifications for various spectrophotometer models.

Manufacturing information

Details about the manufacturer, notified bodies, and declaration of conformity.

LEGAL

GE Novaspec Plus Specifications

General IconGeneral
Wavelength Range190-1100 nm
Spectral Bandwidth2 nm
DetectorSilicon photodiode
Bandwidth2 nm
DisplayLCD
Photometric Accuracy±0.005 A at 1 A
Power Requirements100-240 V, 50/60 Hz
TypeUV-Visible Spectrophotometer
Data OutputRS-232 Interface

Related product manuals