Brief Instruction Card
Operating Instructions
All functions and parameters of the NOVA can be set and recalled on the display at the feed table.
Display indicator for:
1 Developer temperature
2 Fixer temperature
3 Dryer step
4 OK signal
5 SET - key
6 + key
7 - key
8 Value display
9 Status indication / Menu step
1
2
3
5
4
6 7
8 9
DEV
DRY
FIX
DEV
DRY
FIX
DEV
DRY
FIX
DEV
DRY
FIX
DEV
DRY
FIX
DEV
DRY
FIX
DEV
DR
Y
FIX
Setting the process parameters
- Use the (+) and (-) key to scroll the program
- Activate a process setting with the SET key, a dot will appear at the bottom richt side (9)
- The parameter value can now be changed using the (+) or (-) key
- After 10 sec or by pressing SET the changed value is saved
and the display returns to the DEV temperature indication function
Menu Process parameter Range Standard settings
1 Developer temperature 22 - 37 °C 28 °C
2 Fixer temperature 22 - 37 °C 28 °C
3 Dryer step 1 - 6 Step 4
4 Process setting 1.5´ - 12´
5 Developer replenishment 200 - 950 ml/ 900 ml/m²
6 Fixer replenishment 500 - 1250 ml/m² 1200 ml/m²
7 Water replenishment 0 - 15 l/m² 13 l/m² *)
*) The setting of the water feed rate is an optional feature which can only be made when using a Comfort Kit.
Developer temperature (°C)
Fixer temperature (°C)
Dryer step
Process setting
Developer replenishment (x10 ml/m²)
Fixer replenishment (x10 ml/m²)
Water feed rate (l/m²)