4 Command and Query Summary
K0472 Revision A 4-92
:SOUR:PRES:RANG:LOW
Applicability: PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Default:
Query Syntax
:SOURce[x]:PRESsure:RANGe:LOW?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SOUR:LOW?
Function: Queries the lower full-scale of the module [x].
Response: Decimal number in current selected units.
Description
This query gets the lower full-scale of the module [x].
Example 1
TX> :Sour:Pres:Rang:Low?
RX> :SOUR:PRES:RANG:LOW "-1000mbar g"
Example
2
TX> :Sour1:Pres:Rang:Low?
RX> :SOUR:PRES:RANG:LOW "-1000mbar g"