EasyManuals Logo

GE PACE6000 Technical Manual

GE PACE6000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
PACE Series SCPI Manual
K0472 Revision A 4-49
:SENS:PRES
Applicability: PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[:PRESsure]?
where x = 1 or 2 is the module number
(default - 1)
Short form: :SENS?
Function: This query reads the sensor which has been selected by the RANGE
command.
Response: A decimal, pressure reading in the current units.
Description
Queries the pressure reading for the selected sensor in the selected units. The sensor can be
changed see, :SENSe[:PRESsure]:RANGe.
Example
TX> :SENS1:PRES?
RX> :SENS:PRES 3616.9282227
TX> :Sens2:Pres?
RX> :SENS2:PRES 3617.1921387
TX> :SENS2:PRES:RANG?
RX> :SENS2:PRES:RANG "3.50barg"
Currently module 2 sensing range is in gauge mode.
TX> :SENS2:PRES:RANG "4.50bara"
TX> :SENS2:PRES:RANG?
RX> :SENS2:PRES:RANG "4.50bara"
It has been changed to pseudo-absolute mode.
TX> :SENS2:PRES?
RX> :SENS2:PRES 4615.0807447
The reading is also changed to pseudo-absolute mode.
Note:
:SOURe[:PRESsure]:RANGe command does not change ‘SENS?’ reading.
It changes the front-panel display.

Table of Contents

Other manuals for GE PACE6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE6000 and is the answer not in the manual?

GE PACE6000 Specifications

General IconGeneral
BrandGE
ModelPACE6000
CategoryTest Equipment
LanguageEnglish

Related product manuals