5.3 Main menu overview
The main operating menu (main window) shows the wavelength in nanometers, the cell
path length (CL) in millimeters (mm), the absorbance value in absorbance units (AU) and
the mode of operation, STANDBY or RUNNING. The menu is reached from any other
menu by pressing Esc repeatedly.
Setup menu
Setup menu. Setup of analog out, averaging time, serial number etc. Use the arrow keys
to reach setup menu.
Check menu
Check menu. Check internal operating values. Use the arrow keys to reach check menu.
5.4 Setting wavelength
NOTICE
Filter units must be stored in a clean and dustfree place.
NOTICE
Do not run the monitor with an empty filter holder.
The instrument can measure absorbance at wavelengths between 200-700 nm by
changeable filter units. The wavelength currently used is shown in main operating menu.
When no filter is inserted the display shows λ= - - - nm.
Monitor UVis-920 Operating Instructions 29055049 AD 43
5 Operation