Operating Instructions
14 VeriDri™ User’s Manual
7.1 General Specifications (cont)
Operating Temperature
–40° to 60°C (–40° to 140°F)
Storage Temperature
70°C (158
o
F) maximum
Note: The probe should be stored with the plastic cover and desiccant
packet threaded onto the probe. Store it in a cool, dry environment.
Warm-up Time
Meets specified accuracy in 3 minutes
Calibrated Accuracy at 25°C (77°F)
± 2°C (3.6°F) from –65
o
to 10°C (–85
o
to 50
o
F)
± 3°C (5.4°F) from –80
o
to 66°C (–112
o
to 87
o
F)
Repeatability
± 0.5°C (0.9°F) from –65
o
to 10°C (–85
o
to 50
o
F)
± 1.0°C (1.8°F) from –80
o
to –66°C (–112
o
to –87
o
F)
7.2 Electronics
Power
Supply Voltage: 7 to 28 VDC (loop-powered, customer supplied)
Output: 4 to 20 mA
Output Resolution: 0.01 mA
Max. Loop Resistance: R = 50 Ω × (PSV–7),
where PSV = Power Supply Voltage
Example: For a PSV = 24 VDC, R = 50 Ω × (24–7) = 850 Ω
Cable: 2 m, standard (consult factory for custom lengths)