4-28 Section 4-4 - Functional Checks
4-4-9 Calculations 
The Voluson E8 / Voluson E6 system supports calculation packages and application-oriented patient 
Worksheets (Reports) for following applications:
• Abdomen Calculations
• Small Parts Calculations
• Obstetric Calculations
• Cardiology Calculations
• Urology Calculations
• Vascular Calculations
• Gynecology Calculations
• Pediatric Calculations
• Neurology Calculations
• Orthopedics Calculations
NOTE: Confirm that the patient information is correct and the probe and application are selected properly.
NOTE: Different menus are displayed depending on: 
• the currently chosen Application
• the selected Display Mode,
• the selected “Study”, 
• and the settings in the Measure Setup - 
MEASURE & CALC page.
For further details refer to the Voluson E8 / Voluson E6 Basic User Manual.
4-4-9-1 Worksheet (Report) Pages
General remarks to perform Calculations:
• By pressing the CALC
 key on the control panel the Calculation function is switched on.
• Positioning of measurement marks is done with the TRACKBALL
.
• Entering and storage of measuring marks is done with SET
 (right or left trackball key).
• To change measuring marks before completion press CHANGE
 (upper trackball key). 
• Depending on the setting in the Measure Setup, also the FREEZE
 key can be used for confirming 
the last measuring mark of the currently performed measurement.
• The status bar area shows the current function of the trackball.
• To cancel the measurement of the currently selected item, touch CANCEL
 on the Touch Panel.
• To delete the results of the last measured item, touch DELETE LAST
 on the Touch Panel.
• To delete all measurement results of the selected “Study” from the monitor as well as from the 
corresponding Worksheet, touch the CLEAR STUDY
 key on the Touch Panel.
• All measurement results will be automatically included in the corresponding patient worksheet.
• To erase measurement results from the screen, press the CLEAR
 key on the control panel or press 
the DEL
 key on the keyboard.
• To exit from Calculations touch the EXIT
 key on the Touch Panel, press the CALC key or press the 
EXIT
 key on the control panel.
Press the REPORT
 key on the control panel to view the “application dependent” patient worksheet 
pages that contain the results of calculations and measurements. Any stored patient worksheet can be 
edited, printed, transferred, saved in the Archive or sent to DICOM server.