07 - PRESET PROCEDURE SETTINGS
General Densitry Contact
Contact
Contact
Orthodontics
Hygiene
Troughing Yes 0.9 W 1.2 W 1.1 W Initiated Tip
Cuspid Exposure Yes 0.8 W 1.0 W 0.9 W Initiated Tip
Sulcular Debridement Yes 0.4 W 0.5 W 0.4 W Initiated Tip
Decontamination Yes 0.7 W 0.9 W 0.8 W Uninitiated Tip
Incision/Excision Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Aphthous Ulcer No 0.6 W 0.8 W 0.7 W Uninitiated Tip
Class V Gingivoplasty Yes 0.8 W 1.0 W 0.9 W Initiated Tip
Hyperplasia Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Herpetic Ulcer No 0.6 W 0.8 W 0.7 W Uninitiated Tip
Fibroma Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Gingivoplasty Yes 0.9 W 1.2 W 1.1 W Initiated Tip
Gingivectomy Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Molar Exposure Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Aphthous Ulcer No 0.6 W 0.8 W 0.7 W Uninitiated Tip
Operculectomy Yes 1.4 W 1.8 W 1.6 W Initiated Tip
Frenectomy Yes 1.0 W 1.3 W 1.2 W Initiated Tip
Implant Recovery Yes 1.3 W 1.7 W 1.5 W Initiated Tip
Implant Recovery Yes 1.1 W 1.4 W 1.3 W Initiated Tip
Hemostasis Yes 0.8 W 1.0 W 0.9 W Initiated Tip
Frenectomy Yes 1.1 W 1.4 W 1.3 W Initiated Tip
810
810
810
980
980
980
DUAL WAVELENGTH
DUAL WAVELENGTH
DUAL WAVELENGTH
The Gemini 810 + 980 diode laser has 20 preset procedures listed under three categories: General
Dentistry, Orthodontics, and Hygiene. Within each category are the most commonly used procedures
with suggested power settings. Aways use the minimum amount of power necessary to perform
a particular procedure. Manual power adjustment may be necessary depending on patient and
procedural needs.
All Power Settings are shown in Average Power.