032-0207-EN Rev 0
5-2
(Continued)
Table 5-1 Common Symptoms and
Possible Remedies
Symptom Additional Symptoms Possible Cause Remedy
No x-ray output All other functions appear
normal.
SW2 -4 set incorrectly Set SW2-4 from the
ON position to the OFF
position
No time display Mains voltage missing;
power switch OFF or
pluggable terminal strip
not connected
If wiring difficulty is not
easily discerned, have
a qualified electrician
check the power line
F5-F6 fuse blown Contact GENDEX
Technical Support for
further assistance
F1 fuse blown Contact GENDEX
Technical Support for
further assistance
Display flashes "Err0"
and the Ready light is
OFF
Premature release of
the exposure switch
Pressing any key but
the exposure switch on
the Master Control
Touch Panel will clear
the display. Be sure to
press and hold the
exposure button until
the exposure is
finished
Operation may be normal
most of the time, with fault
appearing intermittently
Faulty exposure switch
or wiring
Verify and correct
wiring as necessary, or
replace the exposure
switch
(Table continued on the next page)