5-3
Display flashes "Err1"
and the Ready light is
OFF
Ready light is OFF, other
display functions appear
normal
Mains voltage
configuration jumper
improperly positioned
Re-configure the
jumper per the
instructions on
page 4-
5
Mains voltage out of
range 108 V - 132 V or
198 V - 253 V
Wait until the line
voltage returns to
normal (indicated by
the lamp being steady)
or have a qualified
electrician check the
power line
Display flashes
"Err2"and the Ready
light is OFF
Ready light is OFF;
exposure terminates
normally, but the output
may be below normal
Mains voltage drops
below minimum
required during
exposure, or source
resistance may be too
high
Pressing any key
except the exposure
switch on the Master
Control Touch Panel
will clear the display. If
the problem persists,
have a qualified
electrician check the
power line
No exposure with the
remote exposure
switch
Wrong remote cable
selected from the D800
HK kit
Verify and correct the
wiring as necessary.
Faulty handswitch or
wiring
Mis-wired 110-0160G1
remote cable adapter
kit --customer-supplied
wiring problems
Cooling light is ON Normal operation by
design -- protects and
extends the life of the
tubehead if too many
exposures are made in
too short a period of time
X-ray unit is in the
cooling mode
Wait until the lamp
goes out, indicating
that the tube has
properly cooled down.
(Table continued on the next page)
Table 5-1 Common Symptoms and
Possible Remedies (Continued)
Symptom Additional Symptoms Possible Cause Remedy