8 Calibration and adjustment
207723 rev 7
Gendex Dental Systems 85
8.3 Panoramic calibration
8.3.1 Panoramic geometry calibration
1. Select the program.
2. Press Patient In.
3. Install the double cone calibration tool.
4. Take an exposure.
5. Repeat the calibration until calibration result “passed” is
achieved.
Approved: Laihonen Tuuli 2016-12-02 16:41
Reviewed: Nieminen Timo Antero 2016-12-02 15:31
Approved
See PDM system to determine the status of this document. Printed out: 2017-03-22 10:54:38
D507729, 7
Copyright © 2016 by PaloDEx Group Oy. All rights reserved.