M02 Digital Indicator
28
4.3 Descriptions of Operation Parameters
Code
Default
Description
F1
Null
The first major term of working parameter.
F1.1
OFF
Switch for Auto-Zeroing when power-on,
OFF: disabled ON: enabled
F1.2
0
Zero-tracking Range(0~9d optional).
This parameter is for automatic calibration, disabled
when is set ”0”.
F1.3
1
Motion Detecting Range(1~9d optional)
F1.4
50
Zeroing Range(00%~99% of Maximum capacity)
F1.5
5
Digital filtering parameter: (1-9 as optional)
0: without filtering
9: strongest digital filtering
F1.6
0
Stable filter parameter (the second filter based on the
first filter): (1-9 as optional)
0: without filtering
9: strongest digital filtering
F1.7
0
A/D conversion rate: 120,480,960,15,30,60 as optional